IN THE COURT OF APPEALS OF THE STATE OF IDAHO
Docket No. 38969
IN THE MATTER OF THE SUSPENSION )
OF THE DRIVER’S LICENSE OF LINDA )
LEE HUBBARD. )
LINDA LEE HUBBARD, )
) 2012 Opinion No. 18
Petitioner-Appellant, )
) Filed: April 10, 2012
v. )
) Stephen W. Kenyon, Clerk
STATE OF IDAHO, DEPARTMENT OF )
TRANSPORTATION, )
)
Respondent. )
)
Appeal from the District Court of the Fourth Judicial District, State of Idaho,
Valley County. Hon. Michael R. McLaughlin, District Judge.
District court order affirming decision of hearing officer, affirmed.
Michael G. Pierce, Cascade, for appellant.
Michael Kane & Associates, PLLC; Michael J. Kane, Boise, for respondent.
________________________________________________
LANSING, Judge
Linda Lee Hubbard appeals from the district court’s decision affirming an Idaho
Transportation Department (ITD) hearing officer’s order upholding an administrative suspension
of her driver’s license. We affirm.
I.
BACKGROUND
On September 6, 2010, Hubbard was stopped for a traffic violation and was subsequently
arrested for driving under the influence of alcohol in violation of Idaho Code § 18-8004. She
consented to a breath test, which was administered on a Lifeloc FC20 breath testing instrument.
The test showed breath alcohol concentrations of .113 and .109, and Hubbard’s driver’s license
was administratively suspended. She appealed this suspension to the ITD. Following a hearing,
1
an ITD hearing officer sustained the license suspension. Hubbard filed a petition for review of
the hearing officer’s decision by the district court. The district court affirmed the suspension
order, and Hubbard now appeals. She asserts that her license suspension should be set aside
because her breath test was not conducted in conformity with applicable standards and the testing
equipment was not functioning properly when the test was administered. She argues that the
hearing officer’s findings to the contrary were not supported by the evidence.
II.
ANALYSIS
Idaho Code § 18-8002A(4) directs the ITD to suspend the driver’s license of any driver
who fails an alcohol concentration test. The driver may then request a hearing to contest the
suspension before a hearing officer designated by the ITD. I.C. § 18-8002A(7). In the
administrative hearing, the burden of proof rests upon the driver to prove grounds sufficient to
vacate the suspension. I.C. § 18-8002A(7); In re Suspension of Driver’s License of Gibbar, 143
Idaho 937, 942, 155 P.3d 1176, 1181 (Ct. App. 2006); In re Mahurin, 140 Idaho 656, 658, 99
P.3d 125, 127 (Ct. App. 2004); Kane v. State, Dep’t of Transp., 139 Idaho 586, 590, 83 P.3d 130,
134 (Ct. App. 2003). The hearing officer must uphold the suspension unless the officer finds, by
a preponderance of the evidence, that the driver has shown one of several grounds enumerated in
Section 18-8002A(7) for vacating the suspension. These grounds include that the tests for
alcohol concentration were not conducted by a method that has been approved by the Idaho State
Police (“ISP”) pursuant to Idaho Code § 18-8004(4), and that the testing equipment was not
functioning properly when the test was administered. I.C. § 18-8002A(7)(d).
The hearing officer’s decision is subject to challenge through a petition for judicial
review. I.C. § 18-8002A(8); Gibbar, 143 Idaho at 942, 155 P.3d at 1181; Mahurin, 140 Idaho at
658, 99 P.3d at 127; Kane, 139 Idaho at 589, 83 P.3d at 133. Upon judicial review, a hearing
officer’s decision must be affirmed unless the court determines that the hearing officer’s
findings, inferences, conclusions or decisions are: (a) in violation of constitutional or statutory
provisions; (b) in excess of statutory authority of the agency; (c) made upon unlawful procedure;
(d) not supported by substantial evidence on the record as a whole; or (e) arbitrary, capricious, or
an abuse of discretion. I.C. § 67-5279(3). The reviewing court may not substitute its judgment
for that of the hearing officer as to the weight of the evidence on questions of fact. I.C. § 67-
5279(1). In an appeal from a district court’s decision where the district court was reviewing an
2
agency’s ruling, this Court will review the agency record independently of the district court’s
decision. Castaneda v. Brighton Corp., 130 Idaho 923, 926, 950 P.2d 1262, 1265 (1998).
The ISP has been authorized to adopt methods to analyze breath samples for alcohol
concentration, I.C § 18-8004(4), and to adopt administrative rules governing requirements for
evidentiary testing, including the need to calibrate or check testing equipment. I.C. § 18-
8002A(3). Through this legislative grant of authority, the ISP has approved a number of breath
testing machines for use in Idaho, including the machine at issue here, the Lifeloc FC20. While
the ISP has adopted administrative “Rules Governing Alcohol Testing,” see Idaho
Administrative Code (IDAPA) 11.03.01 et seq., its adopted standards for evidentiary testing and
calibration of equipment are not presented in the body of those rules. Instead, the applicable rule
at the time of Hubbard’s test, former IDAPA 11.03.01.013.03 (3-19-99), 1 stated:
Breath tests shall be administered in conformity with standards established
by the department. Standards shall be developed for each type of breath testing
instrument used in Idaho, and such standards shall be issued in the form of
standard operating procedures and training manuals.
The ISP posts its standard operating procedures and training and reference manuals on its
website, and it amends these documents from time to time. Here we address the administrative
rules, standard operating procedures, and manuals in effect on September 6, 2010, the time of the
administration of Hubbard’s breath tests. We treat those documents as “rules” for purposes of
judicial review because they constitute the only materials by which the ISP has acted upon the
Idaho Code § 18-8002A(3) authorization for the ISP to “prescribe by rule” testing instruments
and methods that are approved by the ISP. In re Schroeder, 147 Idaho 476, 479 n.3, 210 P.3d
584, 587 n.3 (Ct. App. 2009).
The issues raised by Hubbard relate to various “performance verifications” done on the
Lifeloc FC20 instrument used to test her breath. A performance verification is a “verification of
the accuracy of the breath testing instrument utilizing a simulator and a performance verification
solution.” Idaho State Police, 6.0 Idaho Standard Operating Procedure: Breath Alcohol Testing,
Glossary (effective 8/27/10) (“SOP”). A performance verification solution is a “premixed ethyl
alcohol solution used for field performance verifications.” Id. Idaho State Police Forensic
1
Effective April 7, 2011, the rule was amended and renumbered as
IDAPA 11.03.01.014.03.
3
Services tests each lot of verification solution (identified by lot number) to determine its known
alcohol level at the time of testing, which is referred to as a “target value,” and issues a
“certificate of analysis” declaring that target value for that particular lot number of solution. The
certificate of analysis for the verification solution used in this case states a target value of .083.
With respect to the testing instrument at issue here, SOP 5.1.3 specifies that a
performance verification on a Lifeloc FC20 “must be performed within 24 hours, before or after
an evidentiary test to be approved for evidentiary use.” To do this, the operator attaches the
solution to the instrument with a tube, and blows into the tube through a mouthpiece until a result
is obtained. Idaho LIFELOC FC20 Reference Manual (Rev. Effective 8/20/10) (“manual”). A
performance verification consists of two samples, SOP 5.1.2, and is acceptable according to
ISP’s standards if each sample falls within 10 percent above or below the solution’s target value.
SOP 5.1.5.
The agency record before the hearing officer in this case included the arresting officer’s
affidavit, which stated in a conclusory fashion that testing was done in accord with applicable
statutes and the standards and methods adopted by the ISP. Pursuant to a subpoena issued on
Hubbard’s request, a “solution log” for the specific Lifeloc FC20 instrument used here was made
part of the record. The log consists of the arresting officer’s handwritten notations of the results
of performance verifications and evidentiary tests conducted before and after September 6, 2010,
the date of Hubbard’s evidentiary testing. The agency record also included a printout from the
Lifeloc FC20 showing Hubbard’s evidentiary test results.
On appeal, Hubbard first contends that she is entitled to relief from suspension because
her tests for alcohol concentration were not conducted by a method that was approved by the ISP
pursuant to Idaho Code § 18-8004(4). See I.C. § 18-8002A(7)(d). Specifically, she asserts that
the printout of her tests showed that the most recent performance verification prior to her
September 6 breath test was performed on August 27, 2010, and yielded an alcohol concentration
result of .042 for a solution with a target value of .083. She argues that because this result fell
far outside of the acceptable 10 percent variance, the officer violated the prescribed procedures
by conducting her breath test with a machine that had “failed” the most recent performance
verification. We are constrained to disagree. The SOPs do not require that the most recent
performance verification demonstrate that the equipment is operating properly before a breath
test. Rather, SOP 5.1.3 states that a “performance verification solution must be performed within
4
24 hours, before or after an evidentiary test to be approved for evidentiary use.” Under this
standard, an instrument is approved for evidentiary use if it yielded an acceptable performance
verification within twenty-four hours after an evidentiary breath test, even if it had failed the
most recent performance verification preceding the evidentiary test.
Here, although the most recent performance verification before Hubbard’s breath test
showed an unacceptable result, another performance verification conducted within an hour after
her breath test showed results of .081 on each of two samples, which was well within the
acceptable range on the .083 test solution. Therefore, it has not been shown that the officer
failed to comply with standard operating procedures when he conducted Hubbard’s test.
Further, Hubbard’s argument that Lifeloc FC20 should have been taken out of service
after yielding the .042 result on a performance verification on August 27 is incorrect. According
to the SOPs, the only circumstance where a breath testing instrument must be taken out of
service is where it fails three consecutive performance verifications (consisting of six samples)
after certain troubleshooting procedures have been employed and an acceptable result within
verification limits still has not been obtained. SOP 7.1-7.1.5.
Hubbard argues in the alternative that even if the officer complied with the requisite
procedures in conducting her test, her license suspension should be set aside because she
established that the testing equipment was not functioning properly when the test was
administered. This is a basis for setting aside a license suspension under Idaho Code § 18-
8002A(7)(d). Hubbard contends that a malfunction is demonstrated by evidence of a wide
fluctuation in performance verification readings immediately prior to and immediately after her
September 6, 2010, test. She notes that although the log reflects acceptable performance
verification results of .081 and .081 within an hour after her evidentiary test on September 6, the
printout 2 shows an August 27, 2010, performance verification result of .042 on a .083 solution,
while on September 9, according to the officer’s log entries, the same instrument returned
readings of .090 and .089 on the same solution. Hubbard asserts that some of the hearing
officer’s findings that discounted the significance of the .042 and the .090/.089 verification
results are contrary to uncontroverted evidence in the record.
2
The .042 performance verification sample was not entered in the log.
5
We agree that some of the hearing officer’s findings are not supported by the evidence.
The hearing officer said that “Hubbard did not adequately provide proof . . . that the [August 27,
2010] .042 performance verification was the last check prior to Hubbard’s breath test.” The
manual establishes that this finding is wrong. The manual refers to performance verifications as
“wet checks” and states that the Lifeloc FC20 instrument “can store only the most recent [wet
check] result” (emphasis added), and it directs the operator to “[l]og or print the results as the
instrument will overwrite successive wet check results.” The August 27 verification with a .042
result was stored on the instrument at the time of Hubbard’s test, as revealed by a printout from
that instrument. Thus, it was the last performance verification performed on the instrument prior
to Hubbard’s breath test. The hearing officer also found that it is “unknown what simulator
solution lot number was used to perform the September 9, 2010, performance verification
checks.” Contrary to this finding, the solution log shows that the same simulator solution lot
number (9802), which had a certified target value of .083, was used to perform all of the
performance verifications at issue in this case, including the one on September 9, 2010.
These erroneous findings were not directly addressed by the district court because it
concluded that any performance verification results other than the one that was obtained on
September 6, 2010, were not relevant. In substance, the district court held that the ISP’s
acceptable performance verification result on the day of Hubbard’s test established that the
machine was working correctly as a matter of law. We do agree with that analysis. Compliance
with the ISP’s standards for operation of the instrument is not a guarantee that it was operating
correctly. Inadequate or incomplete operating standards and procedures could fail to disclose
when a testing instrument is not functioning correctly. In Gibbar, we interpreted Idaho Code
§ 18-8002A(7)(c) and (d) “as permitting [administrative license suspension] petitioners to
challenge the results of their BAC test by proving that the testing equipment was inaccurate or
was not functioning properly because the State has adopted procedures that do not ensure
accuracy and proper functioning.” Gibbar, 143 Idaho at 947, 155 P.3d at 1186. Similarly, in
State v. Hartwig, 112 Idaho 370, 732 P.2d 339 (Ct. App. 1987), we held that although breath test
results were properly admitted into evidence at a criminal trial because the State showed that the
designated state agency had approved the machine for use and that the machine was operated in
compliance with approved standards, the reliability and performance of the machine is still
subject to challenge. We stated: “If there is evidence that any particular machine has
6
malfunctioned or was designed or operated so as to produce unreliable results, such evidence
would be relevant to both the admissibility and weight of the test results.” Id. at 375, 732 P.2d at
344. See also State v. Ward, 135 Idaho 400, 403-05, 17 P.3d 901, 904-06 (Ct. App. 2001).
Thus, compliance with standards approved by the ISP does not preclude a challenge under Idaho
Code § 18-8002A(7)(d) that the testing equipment was not functioning properly when the test
was administered. Of course, the burden to prove malfunction rests upon the petitioner, as noted
above.
Thus, the question presented is whether Hubbard met her burden to demonstrate that the
testing equipment was malfunctioning when her test was conducted. We conclude that, even
with the hearing officer’s erroneous findings corrected and with all of the evidence properly
considered, Hubbard has not met that burden. The only evidence of any irregularity or anomaly
in the instrument’s performance was the printout that indicated a performance verification
reading of .042 for a solution with a known alcohol concentration of .083 ten days before
Hubbard’s test. There was no testimony as to the significance or possible causes of this
anomaly. For example, it cannot be discerned whether the .042 reading was the result of an
actual, purposeful performance verification test, or some accidental or uncontrolled activation of
the equipment. SOP 7.1.2 identifies a number of factors that could account for an aberrant
verification result, including incorrect use of hoses, not having the instrument properly warmed,
and incorrect blowing technique by the operator. The evidence here does not rule out any of
these possible explanations for the .042 result on the instrument’s printout. In short, while the
.042 performance verification result may raise some question about the instrument’s
performance, standing alone it is insufficient to support a finding that the equipment was
malfunctioning when Hubbard’s test was performed ten days later, particularly in view of the
acceptable performance verification conducted within one hour after her evidentiary test and
several more performance verifications with acceptable results during the month that followed.3
Consequently, notwithstanding certain erroneous findings by the hearing officer, there was no
error in the hearing officer’s ultimate determination that Hubbard had not met her burden to
prove that the testing equipment was not functioning properly when her test was administered.
3
These include the September 9 performance verification that yielded results of .090 and
.089 on the .083 solution. Although these results were near the upper end of the acceptable
10 percent variance, they were still within that range.
7
III.
CONCLUSION
The district court’s decision affirming the hearing officer’s order upholding the
administrative suspension of Hubbard’s driver’s license is affirmed. Costs on appeal to
respondent.
Chief Judge GRATTON and Judge GUTIERREZ CONCUR.
8