In the Missouri Court of Appeals
Eastern District
DIVISION ONE
JOSEPH SCHNEIDER, ) No. ED103011
)
Respondent, ) Appeal from the Circuit Court
) of St. Charles County
vs. ) 1411-CC01176
)
DIRECTOR OF REVENUE, ) Honorable Matthew E.P. Thornhill
)
Appellant. ) FILED: May 10, 2016
OPINION
The Director of Revenue appeals from the judgment of the trial court reinstating the
driving privileges of Joseph Schneider (“Driver”), whose license was suspended for driving
while intoxicated. We reverse.
After a trial de novo pursuant to Section 302.535, 1 the trial court found that the Director
had failed to meet its burden of providing sufficient credible evidence that Driver’s blood alcohol
content (“BAC”) exceeded the legal limit of .08 percent. Crucial to the trial court’s finding was
its contemporaneous decision to exclude from evidence the results from an Alco Sensor IV
breath analyzer showing Driver’s BAC exceeded the legal limit. In the sole point on appeal, the
Director assigns error to the trial court’s decision to exclude the Alco Sensor IV results, a
question of law we review de novo because it hinged on the trial court’s interpretation of an
1
Unless otherwise indicated, all other statutory references are to RSMo (2000) as supplemented.
administrative regulation. See Cortner v. Dir. of Revenue, 408 S.W.3d 789, 792 (Mo. App. E.D.
2013).
To lay a foundation for the admissibility of breath analyzer results, the Director must
establish, inter alia, that the test was performed on equipment and devices approved by the
Department of Health and Senior Services. McGough v. Dir. of Revenue, 462 S.W.3d 459, 462
(Mo. App. E.D. 2015). The breath analyzers must be calibrated using “[c]ompressed ethanol-gas
standard mixtures . . . provided from approved suppliers.” 19 CSR 25-30.051(5) (emphasis here).
The regulation lists four approved suppliers, including Intoximeters, Inc.
In the instant case, the officer calibrating the Alco Sensor IV did so using a mixture
procured from Intoximeters. But because the mixture was actually manufactured by Airgas Mid
America, which is not an approved supplier, the trial court held that Intoximeters was not
“functioning as an approved supplier under the regulations,” and therefore excluded Driver’s
breath analyzer results.
In two analogous cases, we recently had occasion to address whether Intoximeter acts
within the definition of “supplier” when a breath analyzer is calibrated with an ethanol-gas
standard mixture bought from Intoximeter but manufactured by Airgas. See Gallagher v. Dir. of
Revenue, --- S.W.3d ----, 2016 WL 720619 (Mo. App. E.D. 2016); Hiester v. Dir. of Revenue, --
- S.W.3d ----, 2016 WL 720675 (Mo. App. E.D. 2016). We answered that question in the
affirmative, looking to the plain language of “supplier” as distinct from “manufacturer.”
Gallagher, 2016 WL720619, at *2–3. “Intoximeters fits squarely within the plain meaning of
supplier, and further proof of the origins of the gas mixture is not required.” Id. at *3. All that is
required is “that the gas mixture used by law enforcement to maintain the breath analyzer was
provided from one of the approved suppliers listed in the regulation.” Hiester, 2016 WL 720675,
2
at *3. As Driver acknowledged when requesting a stay of briefing in the instant appeal,
Gallagher and Hiester are “dispositive of the only point” on appeal here. 2 Point granted.
We reverse the judgment of the trial court. Because the trial court made no findings on
whether there was probable cause to arrest Driver, and because such probable cause is required
to establish a prima facie case for the suspension of Driver’s license, see McGough v. Dir. of
Revenue, 462 S.W.3d at 464–65 n. 5, we remand the case for further proceedings consistent with
this opinion.
____________________________
Mary K. Hoff, Judge
Robert G. Dowd, Jr., Presiding Judge and Roy L. Richter, Judge, concur.
2
We also note that Driver failed to submit a brief in opposition.